My PowerShell is not charging, what do I do?

By April 26, 2018

Make sure you hear a clicking noise when inserting the Z1 or Z2 into the PowerShell. If it is still not charging disconnect the Z1/Z2, battery, and power supply and reconnect. Proceed to recharge your unit again for 10-12 hours. If it is still not holding a charge it will need to be sent in for a return by the provider you purchased your unit from. Please keep the following information in mind after charging your Power Shell:

  • Disconnect the battery from the PowerShell after a full charge to prevent the battery from slowly draining within a 2-3 day period.
  • The battery needs to be charged for 10-12 hours to ensure a full charge.
  • The bars of battery life on the LCD screen need to be full, and not blinking.
  • The PowerShell battery will not charge while it is operating the Z1/Z2 for treatment. The battery must be charged between treatments.